Welcome
Student Info
Faculty
Research Cores
Contact
Site Map
Search

 

NMR Home
NMR Intro
Policies
Samples
Rates
Documentation
Schedule
NMR Contact
6. TUNING AND SHIMMING

Before any NMR experiment can be performed on a new sample the probe must be tuned and shimmed to obtain the optimum performance of the spectrometer. Both of these operations are critical if the user is to obtain high quality NMR spectra. It is not worth spending weeks preparing a sample if you do not spend the time to tune and shim the probe correctly before carrying out your experiments.
The tuning process matches the resonant frequency of the coil with the sample in place to the input frequency from the spectrometer. Each sample will have a slightly different effect on the resonant frequency whilst different solvents can have very large effects on the resonant frequency. The shimmimg process adjusts the homogeneity of the magnetic field in the vicinity of the sample. This is the most critical and perhaps the most difficult part of setting up your experiment. To the new user it can appear a time consuming and perplexing task however it is worth taking the time and effort to learn how to shim because a poorly shimmed magnet will never give a high quality spectrum.

6.1 SAMPLE VOLUME AND DEPTH

- The magnets have all been shimmed on a sample volume of 500 µl in a 5 mm NMR tube. You should not use a sample volume greater than 500 µl. If you use a sample with a different volume it will be MUCH harder to acheive good shimming within a reasonable time.
- If you do not have sufficient volume for 500 µl of a 1-2 mM sample, you may use a Shigemi tube. These require 320 µl of sample but should only be used when your sample is limited as they are are much harder to shim than the standard volume
- The sample should be in a Wilmad 5 mm NMR tube with the serial no 528 or 535 marked on the outside of the tube.
- The sample tube should be inserted into the correct spinner and the bottom half of the tube should be wiped with a clean tissue to remove dirt and grease.
- Now adjust the position of the tube so that the bottom is at the 62 mm mark in the Varian depth gauge. (Shigemi tubes will need to be lower). Do not touch the bottom of the sample tube with your fingers.
NOTE If your sample is running above room temperature for a period of days, some of the solvent will evaporate from the main body and condense in the upper portion of the tube. This will significantly alter the magnetic field homogeneity during the course of the experiment. If this is likely to be the case for your sample you should discuss the possibility of using a "Matched Susceptibility Plug" to prevent this from occuring.

6.2 REMOVING A PREVIOUS SAMPLE

- Check the "Acquisition Status" window or the "Acquisition Display Unit:".
- If the lock level is NOT zero this means there is a sample in the magnet. If the reading is zero this normally means the probe is empty. However proceed with caution there are times when the lock level will read zero and there is still a sample in the probe.
- Click "Connect" in the Acquisition Window. The window should pop open
- Click on "Eject".
- At this point you should be able to hear the sound of air rushing through the probe. If you cannot hear it WAIT for a few seconds. If you hear a gradual rising tone like somebody whistling, it means there is another sample in the probe and that this is being ejected. Wait until you see the old sample emerge at the top of the magnet.
- Remove the previous sample and place it carefully in the sample holders provided.
- Click "insert" in the Acquisition Window, the air flow should shut off.

6.3 SETTING THE TEMPERATURE

6.3.1 Setting the FTS Controller

The FTS controller consists of two parts: the main control unit which has the LED display and touch pad for adjusting parameters and the compressor unit which cools the input air into the probe if necessary.
- Check the temperature setting on the FTS control unit.
- The Set Point (SP) should be adjusted to 10[ring] C below the temperature required for your sample. Adjust the setting by pressing the s and t arrows on the control unit.
- If the temperature setting on the FTS is to be higher than 25[ring] C, switch off the compressor.
- Wait for a few moments for the temperature to adjust, this should be fairly rapid. If the temperature overshoots and continues to heat or cool in the same direction press the "Run" button twice.

6.3.2 Setting the Sample Temperature

- On the Command line type:
temp=XX where XX is the required temperature
su
- If the temperature was different to the previously set value you should see the value on the display module change. As the temperature approaches the required temperature the rate of change slows and it can take several minutes to regulate the last 1-2 degrees, this is not unusual.
- If your sample is sensitive to high temperatures you should wait a few minutes before putting your sample into the magnet. Otherwise you can proceed immediately.

6.4 INSERTING YOUR SAMPLE

- Open the Acquisition Window and click on "eject"
- Wait until you can hear the sound of the rushing air.
- Place your sample gently into the top of the upper barrel. Make sure that it is free from the sides.
- Click insert in the "Acquisition Window". The sample should now descend into the probe. It will locate with an alarming "Clunk". If it does not then click "eject", wait a few seconds and then click "insert".
- You should now wait 5-20 mins for your sample to equilibrate before you tune and shim the probe. The time you will have to wait will depend on the temperature of your sample before it went into the probe and the temperature you have selected to run the experiments. Clearly if you have taken your sample out of the fridge and are running at 60 [ring]C you will have to wait the full 20 mins. (Go and get a coffee!!). If you are running above room temperature it is advisable to let your sample sit at room temperature for several minutes in order that the dissolved air can come out of the sample.

6.5 TUNING THE 600s

Figure 6.1 shows a schematic layout of one of the spectrometers and a more detailed diagram of the connections on the preamp unit and the magnet/console interface

Figure 6.1 Layout of Preamplifiers and Tuning Unit

6.5.1 Tuning Proton

- Make sure the temperature is set to the required value.
- Check that the following parameters are set.
solvent = `D2O'' If you use a different solvent set this parameter accordingly
tn='H1'
- type "su" in the command window.
- Disconnect the cable (and any filters) from the Probe socket on the 1H preamp (Figure 6.1) and connect it to the Probe socket on the Tune unit. A red light should start to flash.
- Move the cable from the Output socket of the 1H preamp to the Output socket of the tuning unit.
- Press the channel selector button so that it reads 1. The green LCD display will light up
- Set the Attenuation to 8, if the reading is off scale adjust the attenuation to a lower value to get the reading on scale.

 
Figure 6.2 Location of the tuning rods on a Triple Resonance HCN Probe

- On the probe find the rod labeled Proton (Figure 6.2), it usually has a red label
- The rod has two movable parts, the lower smooth part adjusts the "Tune" capacitor and the upper knurled part adjusts the "Match" Capacitor.
- Adjust the Tune capacitor so that the number on the display is a minimum
- Now turn the Match Capacitor a little in a clockwise direction (the number may get bigger)
- Now readjust the Tune to minimize the reading.
- If the new reading is smaller than the previous value, turn the Match in the same direction as before and repeat the tuning process.
- If the new reading is higher than the original reading, turn the Match back in the opposite direction and readjust the Tune to minimize the display,
- Keep adjusting the Match and Tune in this way until the reading reaches a minimum. It should be possible to get the meter reading down to 1-2 on all probes with the attenuation set to 8.
- When you have finished set the channel selector to 0 and move the cables back to their original positions

6.5.2 Tuning the First Decoupler Channel

Most modern NMR experiments use samples that are labeled with C13 and/or N15 and it is necessary to apply pulses or decoupling to the second channel. The larger spectral widths of C13 experiments place a greater demand on the power requirements for applying RF pulses. The first decoupler channel is routed through the Magnet-Console interface. The presence of relays and filters within the interface leads to a loss in 2-3 dB of power compared to the second decoupler channel which is routed directly from the power amplifier into the probe. Consequently the spectrometers are normally configured to apply N15 pulses on the first decoupler channel and C13 pulses on the second decoupler channel. If you wish to change this you can do so.
- Set tn='H1'
- Set dn='N15' or 'C13'
- type su
- Tune the Proton channel as above
- Connect the Cable from the B-Band Out socket on the back of the Magnet/Console Interface to the "Probe" input of the tuning unit.
- Connect the output cable from the X-Preamplifier to the output of the tune unit
- Set the channel selector switch to 2
- Adjust the 'C13' or 'N15' rods (as appropriate) as before

6.5.3 Tuning the Second Decoupler Channel for C13

Tuning the second decoupler channel is similar to tuning the first decoupler channel except that intially the cables are connected differently:
- Set tn=`H1'
- Set dn=`N15'
- Set dn2=`C13'
- Type su
- Tune the Proton channel and the N15 channel as described in sections 6.5.1 and 6.5.2 above
- Locate the cable coming from the C13 channel on the probe. It should be connected to a large cylindrical filter. The other end should be connected to a thick black cable labeled 2ND DEC or DEC2
- Disconnect the thick cable and connect a one end of a thin cable to the cylindrical filter and connect the other end to the Probe socket on the tune unit
- Connect the Output from the X-Preamplifier to the Output of the Tune unit
- Set the channel selector switch to 3 and tune as before
- When you have finished, reconnect the thick cable labeled DEC2. DO NOT confuse it with the cable labelled DEC3. Replace all other cables in their original positions

6.5.4 Tuning X Nuclei for Direct Observe

- Repeat the above procedure except for the following changes
- Set tn='N15', 'C13' or 'P31' etc.
- Set dn='H1'
- Set the channel selector to 1 and tune the X nucleus
- When you have finished connect the Probe cable to the Probe socket on the X Preamp
- Tune the 1H channel as before except set the channel selector to 2
- Connect the 1H Probe Cable to the H-Band Out socket on the Magnet/Console Interface

6.6 TUNING ON THE 300 AND 400

Tuning on the 300 and 400 is essentially the same as the 600s except you only need to reconnect one cable to perform any tuning. This is the cable that comes from the Probe and goes directly into the back of the Magnet/Console Interface. You disconnect this cable and connect it to the Probe input on the Tune interface which is located on the front of the pre-amplifier housing and not the magnet leg.
The 300 and 400 are equipped with different hardware which can only produce 1H frequencies on the 2nd channel. Any other frequency must be produced from the 1st transmitter. The spectrometer checks the setup for each experiment and automatically switches the 1H to the second channel if necessary. You must be aware of this change if you run experiments with 1H and X nuclei set.

6.6.1 Tuning for 1H Only Experiments

- Set tn and dn ='H1'
- Set the Channel Selector to 1

6.6.2 Tuning for any Other Experiment

- For 1H observe/ X decouple experiments set tn='H1' and dn='X'
- For X observe/ 1H decouple experiments set tn='X' and dn='1H'
- In both cases, connect the Proton cable to the tuning meter and set the channel selector to 2 and tune the proton as before
- Then connect the X cable to the tuning meter and set the channel selector to 1 and tune as before

6.7 LOCKING

The lock circuit is used to maintain a stable magnetic field. It uses the deuterium signal of the solvent as a fixed reference frequency. The lock circuitry detects changes in the magnetic field and adjusts the field strength accordingly to maintain the deuterium signal at the reference frequency. The exact frequency of the deuterium signal is different for each solvent but the spectrometer uses a list of the deuterium frequencies stored in one of the configuration files to adjust its parameters to accommodate these differences.
- Check the setting of the "solvent" parameter. This is normally set to D2O, however on some occasions other users will use different solvents e.g. DMSO.
- Check that the temperature is set correctly in this experiment.
- Enter the command "su"
- Open the Acquisition window
- Click on the "Lock" Button
- In the pop-up window click the lock "OFF" button
- Change the "Lock Power": For D2O set the Lock Power to 25, for H2O set the lock power to 36
- Adjust the "Lock Gain" so that you can see the lock signal
- Adjust the value of Z0 to get maximum intensity. The final signal should have no modulation and only a small decay
- Click the lock "ON" button
- Adjust the "Lock Phase" to maximize the lock level

6.8 SHIMMING

6.8.1 Introduction

Shimming is the process of adjusting the magnetic field to produce a homogeneous field throughout the sample volume. Good shimming is essential for all experiments, particularly those that use selective pulses and gradients for suppression of strong unwanted solvent resonances. The number of shims on each magnet varies from 13 on the 300 to 39 on the 600s. They can be classified into two major groups: The Axial shims (also called the Z shims) are those that contain only Z components, i.e. Z1, Z2, Z3, Z4 etc.. The Radial shims (also known as non-spinning shims) are those that contain X and/or Y components, they may also contain Z components e.g. X, Y, XZ, YZ, XZ2, YZ2 etc.

There are two methods for shimming your sample, the first is the manual way by hand. The other is to use Gradient Shimming. This latter method uses gradients to generate maps of the effects of each shim on your sample and applies a mathematical correction to the shims. It is a very powerful and rapid technique that can acheive amazing results very quickly. The process for Gradient Shimming is discussed in detail below.

The shims have initially been adjusted for the best sample homogeneity by the Varian Engineers during the installation of the machines. However, each time you put a sample in the magnet you may have to adjust the following shims: Z1-Z5, X, Y, XZ, YZ. The Z shims are the most sensitive with Z1 being the most sensitive, but Z4 and Z5 have the most complex effects on the sample. The higher order shims are normally adjusted using a sample of Chloroform in Acetone which has a line width of less than 0.2 Hz. Protein samples typically have line widths 100 times greater than this and it is difficult to observe any effects on the

6.8.2 Manually Shimmimg Your Sample

In practice shims are normally adjusted in pairs. Each pair of shims is adjusted in an iterative fashion to produce the maximum signal possible before the next pair is adjusted. When all sets have been adjusted the whole process is repeated. The following scheme suggests one method for iterating the shims.

Step 1 - Adjust Z1 and Z2
Step 2 - Maximize Z3 and then readjust Z1 and Z2

      - Repeat the above until there is no further increase in the lock level
Step 3 - Move Z4 by 128 and repeat Step 1
    - If the lock signal increases repeat the above step
    - If the lock level is lower adjust Z4 in the opposite direction and repeat step 1
    - Repeat the above steps until no there is no further increase in the lock level
    - Periodically check the adjustment of the lock phase as large changes in Z4 can induce significant changes in the phase of the lock signal
Step 4 - Adjust X and Y
         
            - Adjust XZ and YZ and then readjust X and Y
            - Then repeat step 1-3

Step 5 - Repeat Steps 1 to 4 until there is no further increase in the lock level.

If you have problems shimmimg your sample you can put in a test sample of 99.99% D2O with the same sample volume as your sample and shim on this to get a good lineshape. Then replace it with your own sample making sure they are set at the same level in the depth guage. You will have to adjust Z1, Z2 and Z3 but the remaining shims should be very close to the required values. If their are significant differences in the sample depth you will have to apply large corrections to the Z4 shim. This cannot be done on a 90% H2O sample and must be done by using a D2O sample.

6.8.3 Storing and Retreiving Shim Values

If it takes you a long time to shim your sample and you intend to use this sample for other NMR experiments in the future you can store the shim values for that sample. There are two ways to do this but we recommend that you save the shims in a 1D data file. To do this set up any of the 1D experiments outlined below and set nt=1, run the single scan experiment. Now change to the shim directory under vnmrsys and save the file with a meaningful name e.g. sry_m64i_122595_61mm. To retreive these shim values load the file and then enter the following commands:
- load=`y' su

NOTE The shim values are only valid for the machine and probe for which they were adjusted. If you use a different machine or a different probe you must reshim your sample. Furthermore your shim values are only valid if the sample is exactly the same next time you use it. Even then you will have to readjust the Z1-Z3 shims.

6.9 GRADIENT SHIMMING

Gradient shimming is the best way to acheive good line shape. There are two distinct stages to the gradient shimming process the first is to Generate a Shim Map for your sample. The second is to perform the actual gradient shimming. In either case it is critical to have well adjusted radial shims (X,YXZ and YZ etc...).

6.9.1 Generating Shim Maps

The first step in generating a shim map is to optimise Z1, Z2, X,Y XZ and YZ shims manually. This is critical if you want to produce a good map on the first pass. You should also check the other higher order radial shims. However if you do not see any significant change in the lock level you should put them back to their initial settings.

Having made an initial adjustment to the shims we can now set up the mapping parameters. Enter the command "gmapsys".  You should now get a new set of menu buttons on the command line.

There are X stages to Generating a Shim Map. The first is to set up the  parameters such as pulse width, offset and gradient strengths, then set the optimal window for generating the shim map and finally generate the shim map itself.


Previous Top Contents Next

 


The University of Colorado Health Sciences Center is committed to equal opportunity and affirmative action.



Website Manager